INSTRUCTIONS FOR USE REASTAIN Quick-Diff is an optimised colour and fixative composition for the differential staining of cellular elements in peripheral blood smears. The fast 3- minute staining procedure and the ready-to-use solutions of the REASTAIN Quick-Diff KIT make this method ideal for the emergency units. The kit includes three 500 ml vials; REASTAIN Quick-Diff BLUE, REASTAIN Quick- Diff RED and REASTAIN Quick-Diff FIX. The result is in differential staining of the white cells upon their cytochemical features. Reagents needed for the differential staining Product Size code Reastain Quick-Diff Kit 3 x 500 ml 102164 Phosphate buffer 67,0 mmol/l 100 ml 140105 pH 6,8 500 ml 170105 10 l 110105 For professional use only.
Reference intervals Microscopic differential count of blood white cells (adults): Parameter %, fraction Neutrophils 40-75 Eosinophils 1-5 Basophils <1 Monocytes 1-10 Lymphocytes 20-45 Samples EDTA blood or skin puncture sample.
Preparation of the samples See Dacie and Lewis (1995). Preparation of applicable and good quality specimens requires special attention.
Contents of the staining reagents: Reastain Quick-Diff Blue Azur II 0,09 % Glycerol 5 % Sodiumazide <0,1 % Reastain Quick-Diff Fix Methanol 100 % Methyleneblue <0,01 % Reastain Quick-Diff Red Eosin Y 0,12 % Sodiumazide <0,1 %
Phosphate buffer Potassium dihydrogen phosphate/ disodium hydrogen phosphate x 2H2O 67,0 mmol/l Storage Reastain Quick-Diff Kit: protected from light at 2?25 °C. Unopened reagents may be used until the expiry date on the label. Phosphate buffer: at 2?8 °C. Unopened reagents may be used until the expiry date on the label.
Applicability The reagents can be applied on manual staining procedures.
STAINING INSTRUCTIONS Preparation of buffered water Dilution ratio 1:20, eg. 10 ml phosphate buffer + 190 ml deionised or distilled water. Staining 1. Pour the solutions into the separated staining chambers. 2. Dip the air-dried slides into the REASTAIN Quick-Diff FIX solution (1sec) 5 - 10 times. 3. Dip the slides into the REASTAIN Quick-Diff RED solution (1sec) 3 - 15 times. 4. Dip the slides into the REASTAIN Quick-Diff BLUE solution (1sec) 3 - 15 times. 5. Rinse the slides in the buffered water solution, and air-dry. 6. Mount the slides with a coverslip if longer storage is required. The staining result can be modified by the dipping procedures. For example, dipping the slides into the REASTAIN Quick-Diff RED solution more frequently than into the REASTAIN Quick-Diff BLUE solution can increase staining intensity of the eosinophilic cells. The user should establish the most appropriate procedure for the different purposes by altering the dipping procedure.
One possible dipping procedure established by us is as follows: 1. REASTAIN Quick-Diff FIX 10 times 1 second each 2. REASTAIN Quick-Diff RED 6 times 1 second each 3. REASTAIN Quick-Diff BLUE 10 times 1 second each
Attentions! Use fresh and non-expired solutions. The staining solutions in staining chambers should be changed periodically to ensure optimal staining. Prevent REASTAIN Quick-Diff FIX from evaporation. Contains methanol.
Sources of errors Irregular distribution of the blood smear on a glass slide may result in an erroneous cell counts. Alcohols used for wiping the skin may cause hemolysis and artifacts. Do not let the specimens dry at any stage of the staining procedure. Wash properly to avoid dye artifacts. Buffered water is strongly recommended for washing. Staining result is dependent on pH. Alkaline pH increases blue and acidic pH pink or reddish tinge in the stained specimen. Test performance When stained correctly all relevant cell types are stained individually and are identifiable by the professional.
Warnings Reastain Quick-Diff Fix: Highly flammable. Toxic by inhalation, in contact with skin and if swallowed. Toxic: danger of very serious irreversible effects through inhalation, in contact with skin and if swallowed. Keep container tightly closed. Keep away from sources of ignition - no smoking. Wear suitable protective clothing and gloves. In case of accident or if you feel sick, seek medical advice immediately (show the label if possible).
Reference Dacie J. and Lewis S. Practical haematology. Churchill Livingstone, London, 1995. |